A visual odometry base-tracking system for intraoperative C-arm guidance
- PDF / 3,627,592 Bytes
- 13 Pages / 595.276 x 790.866 pts Page_size
- 38 Downloads / 166 Views
ORIGINAL ARTICLE
A visual odometry base‑tracking system for intraoperative C‑arm guidance Luke Haliburton1 · Hooman Esfandiari2 · Pierre Guy3 · Carolyn Anglin4,5,6 · Antony Hodgson1 Received: 13 February 2020 / Accepted: 6 July 2020 © CARS 2020
Abstract Purpose C-arms are portable X-ray devices used to generate radiographic images in orthopedic surgical procedures. Evidence suggests that scouting images, which are used to aid in C-arm positioning, result in increased operation time and excess radiation exposure. C-arms are also primarily used qualitatively to view images, with limited quantitative functionality. Various techniques have been proposed to improve positioning, reduce radiation exposure, and provide quantitative measuring tools, all of which require accurate C-arm position tracking. While external stereo camera systems can be used for this purpose, they are typically considered too obtrusive. This paper therefore presents the development and verification of a low-profile, real-time C-arm base-tracking system using computer vision techniques. Methods The proposed tracking system, called OPTIX (On-board Position Tracking for Intraoperative X-rays), uses a single downward-facing camera mounted to the base of a C-arm. Relative motion tracking and absolute position recovery algorithms were implemented to track motion using the visual texture in operating room floors. The accuracy of the system was evaluated in a simulated operating room mounted on a real C-arm. Results The relative tracking algorithm measured relative translation position changes with errors of less than 0.75% of the total distance travelled, and orientation with errors below 5% of the cumulative rotation. With an error-correction step incorporated, OPTIX achieved C-arm repositioning with translation errors of less than 1.10 ± 0.07 mm and rotation errors of less than 0.17 ± 0.02◦ . A display based on the OPTIX measurements enabled consistent C-arm repositioning within 5 mm of a previously stored reference position. Conclusion The system achieved clinically relevant accuracies and could result in a reduced need for scout images when re-acquiring a previous position. We believe that, if implemented in an operating room, OPTIX has the potential to reduce both operating time and harmful radiation exposure to patients and surgical staff. Keywords C-arm · Computer vision · Tracked C-arm · Radiation · Orthopedic · Position tracking
Introduction * Luke Haliburton [email protected] 1
Department of Mechanical Engineering, University of British Columbia, Vancouver, BC, Canada
2
Surgical Technologies Laboratory, Biomedical Engineering, University of British Columbia, Vancouver, BC, Canada
3
Department of Orthopaedics, University of British Columbia, Vancouver, BC, Canada
4
Department of Civil Engineering, University of Calgary, Calgary, AB, Canada
5
Biomedical Engineering, University of Calgary, Calgary, AB, Canada
6
McCaig Institute for Bone and Joint Health, Calgary, AB, Canada
C-arms are mobile fluoroscopy mac