Comparative Study for Measurement of Residual Leucocytes in Leucodepleted Red Blood Cells by Two Different Methods

  • PDF / 216,668 Bytes
  • 5 Pages / 595.276 x 790.866 pts Page_size
  • 52 Downloads / 285 Views

DOWNLOAD

REPORT


ORIGINAL ARTICLE

Comparative Study for Measurement of Residual Leucocytes in Leucodepleted Red Blood Cells by Two Different Methods Prashant Pandey1 Ajay Shanker1



Amit Pande1 • Divya Setya1 • Praveen Kumar1



Received: 17 April 2020 / Accepted: 20 July 2020 Ó Indian Society of Hematology and Blood Transfusion 2020

Abstract Estimation of residual leukocytes in blood components after leukodepletion is crucial for assessment of quality. Flow cytometry (FC) and Nageotte hemocytometer are the most widely accepted methods for counting residual white blood cells (rWBCs) in leucocyte-depleted (LD) blood components. The objective of this study was to compare use of Nageotte counting chamber and FC for quality control of leukodepleted red cell units. A prospective, observational study was conducted in the department of Transfusion Medicine. A total of 80 whole blood donations from healthy donors were subjected to testing by FC and Nageotte hemocytometer for estimation of rWBC in duplicate. Additionally, ten personnel attempted a survey for ease of use of FC. Number of rWBC detected by flow cytometer were between 1 WBC/lL and 28 WBCs/lL whereas that detected by Nageotte’s chamber were between 0 WBC/ lL (lowest) and 5 WBCs/lL. Coefficient of variation was found to be 87.36% by Nageotte hemocytometer method and 43.26% by FC. Linear regression analysis

& Prashant Pandey [email protected] Amit Pande [email protected] Divya Setya [email protected] Praveen Kumar [email protected] Ajay Shanker [email protected] 1

Department of Transfusion Medicine, Histocompatibility and Molecular Biology, Jaypee Hospital, Sector-128, Noida 201304, India

did not show any correlation (R-squared = 0.01, p = 0.13) between the two methods which signifies that the two methods cannot be used interchangeably. Pearson’s correlation coefficient showed a weak relation between results obtained by the two methods. Inter-observer variation was found to be significant with use of Nageotte hemocytometry. Survey for ease of use of FC indicated acceptability of FC with favorable scores. Flow cytometric technique provides a reproducible and objective tool for counting rWBC in leukodepleted blood components compared with the Nageotte hemocytometer. Keywords Leukoreduction  Leukodepletion  Filtration  Flowcytometry  True count beads  Nageotte

Introduction Blood transfusion is a very common medical procedure and it carries an inherent risk of infectious and noninfectious adverse events [1, 2]. To enhance safety of blood transfusion, many policies and procedures exist in any patient care setting. One of them is leukodepletion (LD) of blood components, which is used to remove the leukocytes that are known to cause non hemolytic febrile transfusion reaction (NHFTR), alloimmunization and refractoriness, CMV transmission and immunomodulation among others [3, 4]. It is now established that reducing leukocytes from blood components would result in reducing these adverse reactions [5, 6]. Methods for LD can be classified ba