Examining trabecular morphology and chemical composition of peri-scaffold osseointegrated bone

  • PDF / 1,623,385 Bytes
  • 9 Pages / 595.276 x 790.866 pts Page_size
  • 68 Downloads / 158 Views

DOWNLOAD

REPORT


(2020) 15:406

RESEARCH ARTICLE

Open Access

Examining trabecular morphology and chemical composition of peri-scaffold osseointegrated bone Linwei Lyu1,2 , Shicai Yang3, Ye Jing1,2, Chunqiu Zhang1,2* and Jikun Wang4

Abstract Background: Porous titanium alloy scaffold fabricated by 3D printing technology could induce osseointegration well to repair bone defect during early postoperative period. However, trabecular histomorphological features and chemical compositions of ingrowth bone in the long term after surgery still lacked in-depth research. Methods: Fourteen New Zealand rabbits were divided into two groups (7 rabbits in surgery group and 7 rabbits in control group). A 3D-printed porous titanium alloy scaffold was implanted into right femoral condyle of each rabbit in the surgery group. Preload was produced at the surface between bone tissue and scaffold through interference assembly during implantation process. Rabbits in the control group were feed free. All rabbits were sacrificed to extract femoral condyles at week 12 after surgery. All right femoral condyles were performed micro-CT scanning to test bone mineral density (BMD) and trabecular histomorphological parameters, including bone volume fraction (BV/TV), bone surface/volume ratio (BS/BV), bone surface density (BS/TV), structure model index (SMI), trabecular thickness (Tb.Th), trabecular number (Tb.N), trabecular separation (Tb.Sp), porosity (PO), connectivity density (Conn.Dn), and degree of anisotropy (DA). Scanning electron microscope was used to observe osteogenesis peri-scaffold. Fourier transform infrared spectroscopy (FTIR) scanning was performed to analyze chemical compositions of peri-scaffold trabeculae. All trabecular morphological parameters and BMDs were statistically analyzed between surgery group and control group. Results: The pores of scaffold were filled with ingrowth bone tissues after 12 weeks osseointegration. However, the mean BMD peri-scaffold in surgery group was 800 ± 20 mg/cm3, which was 18.37% lower than that in the control group. There was a significant decrease in BV/TV, Tb.N, and BS/TV, and there was a significant increase in Tb.Sp and PO between the surgery group and control group (p < 0.05). There were no significant differences in Tb.Th, SMI, Conn.Dn, BS/BV, and DA. Although ingrowth of bone tissue was very effective, some fragmented connective tissues were still found instead of bone tissues on the partial beams of scaffolds through SEM images. It was found from FTIR that there was no significant hydroxyapatite peak signal in surgery group. Collagen in the control group mainly existed as crosslink structure, while non-cross-link structure in the surgery group. (Continued on next page)

* Correspondence: [email protected]; [email protected] 1 Tianjin Key Laboratory for Advanced Mechatronic System Design and Intelligent Control, Tianjin University of Technology, No. 391 Binshui Xidao, Xiqing District, Tianjin, China 2 National Demonstration Center for Experimental Mechanical and Electrical Engineering Education (Tian