Patient-specific drill template for C2 transoral pedicle insertion in complete reduction of atlantoaxial dislocation: ca

  • PDF / 3,427,315 Bytes
  • 8 Pages / 595.276 x 790.866 pts Page_size
  • 53 Downloads / 177 Views

DOWNLOAD

REPORT


(2019) 14:141

RESEARCH ARTICLE

Open Access

Patient-specific drill template for C2 transoral pedicle insertion in complete reduction of atlantoaxial dislocation: cadaveric efficacy and accuracy assessments Lijun Lin1†, Meisong Zhu1†, Peng Peng2, Xintao Zhang3, Xiaoqi Zhou1 and Jianyi Li2,4*

Abstract Background: The transoral atlantoaxial reduction plate (TARP) is an effective advance in the treatment of atlantoaxial dislocation (AAD) and can enable the performance of anterior atlantoaxial release, reduction, decompression, and internal fixation in a one-stage operation. However, accurate transoral C2 pedicle insertion (C2TOPI) remains a challenge. The aim of this study is to develop a grouped patient-specific drill template (PDT) specifically for AAD with complete reduction and, furthermore, to compare its efficacy and accuracy in facilitating C2TOPI. Methods: After CT scanning, ten cadaveric C2 specimens were randomly assigned to two groups (the PDT and freehand group). A grouped PDT specifically for AAD with complete reduction was designed and manufactured. C2TOPI was performed using the PDT or the fluoroscopy-guided freehand technique. Postoperative CT scans were subsequently performed to analyze the deviations at the centroid of the cross section at the midpoint of the pedicle. Screw position grades were also assessed in both groups. Results: Compared to the freehand group, the PDT group had a significantly shorter surgery time (p < 0.001). Significant differences between the two groups were observed in the absolute value of the deviations at the centroid of the pedicle on either the axial or sagittal planes (p < 0.05). No significant difference was found in the screw positions between the two groups (p > 0.05); however, two unacceptable breaches (20%) occurred in the freehand group. Conclusion: A specifically designed PDT could provide an accurate and easy-to-apply method for C2TOPI in AAD with complete reduction. Keywords: Patient-specific drill template, Transoral atlantoaxial reduction plate, Atlantoaxial dislocation, Pedicle fixation

* Correspondence: [email protected] † Lijun Lin and Meisong Zhu contributed equally to this work. 2 Department of Anatomy, Guangdong Provincial Key Laboratory of Medical Biomechanics, School of Basic Medical Sciences, Southern Medical University, 1063 Shatai Nan Road, Baiyun District, Guangzhou, Guangdong, China 4 Nanhai Hospital, Southern Medical University, 28 Liguan Road, Lishui Town, Nanhai District, Foshan, Guangdong, China Full list of author information is available at the end of the article © The Author(s). 2019 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver