Post-operative monitoring of free flaps using a low-cost thermal camera: a pilot study

  • PDF / 860,954 Bytes
  • 8 Pages / 595.276 x 790.866 pts Page_size
  • 8 Downloads / 171 Views

DOWNLOAD

REPORT


ORIGINAL PAPER

Post-operative monitoring of free flaps using a low-cost thermal camera: a pilot study S. Hummelink 1 & A. S. Kruit 1 & A. R. W. van Vlaenderen 1 & M. J. M. Schreinemachers 1 & W. Steenbergen 2 & D. J. O. Ulrich 1 Received: 24 July 2019 / Accepted: 2 March 2020 # The Author(s) 2020

Abstract Background Careful post-operative monitoring of free flaps is important in flap survival; immediate action increases flap salvage rate. Although various methods are available, room for improvement remains. Thermal cameras have proven their value in medicine and are nowadays readily available at low costs. The objective of this study was to evaluate the potential of an affordable infrared thermal camera and software in the detection of failing free flaps during post-operative monitoring. Methods Free myocutaneous rectus abdominis flaps were harvested in 16 female landrace pigs and replanted after several hours of storage. All flaps were assessed with indocyanine green fluorescence angiography as well as hourly clinical assessment of skin colour, turgor and capillary refill. Furthermore, thermal photographs were taken simultaneously with the FLIR One thermal camera smartphone module. These photographs were processed in MATLAB and evaluated on their additional value as an indicator for flap failure. Results Out of 16 flaps, three flaps failed due to arterial failure and one flap developed venous congestion. The mean flap temperature compared to adjacent control skin proved to be most indicative for flap failure. All unsuccessful flaps showed lower temperatures after failure compared to the uncompromised free flaps. Conclusions An affordable thermal camera module can potentially contribute to post-operative free flap monitoring. Vascular compromise in free flaps can be distinguished by investigating relative temperature differences between the flap and reference skin. Until the FLIR One camera has been extensively investigated in a human population, it should be used in conjunction with conventional monitoring techniques. Level of evidence: Level IV, diagnostic study Keywords Thermal camera . Flap monitoring . Innovation . Flap surgery . Smartphone

Introduction Over the last decades, free tissue flaps have been extensively used for the reconstruction or coverage of large defects following trauma, burns, infection and tumour extirpation [1]. During this procedure, the flap is completely * S. Hummelink [email protected] 1

Department of Plastic Surgery, Radboud University Medical Center, Nijmegen, The Netherlands

2

Biomedical Photonic Imaging, Technical Medical Centre, University of Twente, Enschede, The Netherlands

separated from the donor site, and the blood supply is microsurgically restored by vascular anastomoses at the recipient site. Although the overall success rate of free flap transfer is high, 5–25% of transferred flaps require surgical revision due to circulatory compromise [2]. The majority of failures are caused by venous (54–57%) and arterial thrombosis (20–43%), mostly occurring within