A novel amplitude binning strategy to handle irregular breathing during 4DMRI acquisition: improved imaging for radiothe
- PDF / 3,104,331 Bytes
- 14 Pages / 595.276 x 790.866 pts Page_size
- 115 Downloads / 151 Views
RESEARCH
Open Access
A novel amplitude binning strategy to handle irregular breathing during 4DMRI acquisition: improved imaging for radiotherapy purposes Z. van Kesteren1* , A. van der Horst1, O. J. Gurney-Champion1,2,4, I. Bones1, D. Tekelenburg1, T. Alderliesten1, G. van Tienhoven1, R. Klaassen3, H. W. M. van Laarhoven3 and A. Bel1
Abstract Background: For radiotherapy of abdominal cancer, four-dimensional magnetic resonance imaging (4DMRI) is desirable for tumor definition and the assessment of tumor and organ motion. However, irregular breathing gives rise to image artifacts. We developed a outlier rejection strategy resulting in a 4DMRI with reduced image artifacts in the presence of irregular breathing. Methods: We obtained 2D T2-weighted single-shot turbo spin echo images, with an interleaved 1D navigator acquisition to obtain the respiratory signal during free breathing imaging in 2 patients and 12 healthy volunteers. Prior to binning, upper and lower inclusion thresholds were chosen such that 95% of the acquired images were included, while minimizing the distance between the thresholds (inclusion range (IR)). We compared our strategy (Min95) with three commonly applied strategies: phase binning with all images included (Phase), amplitude binning with all images included (MaxIE), and amplitude binning with the thresholds set as the mean end-inhale and mean end-exhale diaphragm positions (MeanIE). We compared 4DMRI quality based on: Data included (DI); percentage of images remaining after outlier rejection. Reconstruction completeness (RC); percentage of bin-slice combinations containing at least one image after
binning. Intra-bin variation (IBV); interquartile range of the diaphragm position within the bin-slice combination,
averaged over three central slices and ten respiratory bins. IR. Image smoothness (S); quantified by fitting a parabola to the diaphragm profile in a sagittal plane of the
reconstructed 4DMRI. A two-sided Wilcoxon’s signed-rank test was used to test for significance in differences between the Min95 strategy and the Phase, MaxIE, and MeanIE strategies. (Continued on next page)
* Correspondence: [email protected] 1 Department of Radiation Oncology, Amsterdam UMC, University of Amsterdam, Meibergdreef 9, 1105AZ Amsterdam, The Netherlands Full list of author information is available at the end of the article © The Author(s). 2019 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.