A silver nanowire-based flexible pressure sensor to measure the non-nutritive sucking power of neonates
- PDF / 1,812,689 Bytes
- 8 Pages / 595.276 x 790.866 pts Page_size
- 80 Downloads / 163 Views
pen Access
LETTER
A silver nanowire‑based flexible pressure sensor to measure the non‑nutritive sucking power of neonates Jean G. de Oliveira, Tausif Muhammad and Sohee Kim*
Abstract Preterm infants are prone to have higher risks of morbidity, disability and developmental delay compared to term infants. The primitive reflexes, inborn behaviors found in early life development, are shown to be a good tool to assess the integrity of the central nervous system of infants and to predict potential malfunctions. Among these reflexes, the non-nutritive sucking reflex plays an important role in indicating congenital abnormalities in brain development and feeding readiness, especially for premature infants. Conventionally, pediatricians evaluate the oral sucking power qualitatively based on their experiences, by using a gloved finger put inside the infant’s mouth. Thus, more quantitative solutions to assess the sucking power of preterm infants are necessary to support healthcare professionals in their evaluation procedures. Here, we developed a silver nanowire (AgNW)-based flexible pressure sensor to measure the non-nutritive sucking power of infants. The flexible sensor was fabricated using silver nanowires deposited on polydimethylsiloxane (PDMS) in a sandwich-like structure. The sensor based on the principle of strain gauge was attached to a ring-shaped connecting module, and then to a pacifier. The negative sucking pressure exerted by the infant deformed the sensor membrane, causing its electrical resistance to change without any contact between the infant’s mouth and the sensing element. The fabricated sensor was characterized and optimized to achieve both the suitable sensitivity and stability. Thanks to the excellent long-term electro-mechanical stability and high sensitivity, the developed sensor is expected to provide the means to quantitatively assess the non-nutritive sucking of infants, with a portable, low-cost, non-invasive and light-weight solution. Keywords: Premature infant, Non-nutritive sucking, Pressure sensor, Flexible sensor, Silver nanowires Introduction The World Health Organization (WHO) defines preterm birth as any birth before 37 completed weeks of gestation, or fewer than 259 days since the first day of the woman’s last menstrual period [1]. This definition can be sub-divided based on gestational age (GA), as moderate or late preterm (32 to 37 completed weeks of gestation); very preterm (28 to 32 weeks of gestation); extremely preterm (less than 28 completed weeks of gestation). In addition, the WHO estimates that 15 million preterm *Correspondence: [email protected] Department of Robotics Engineering, Daegu Gyeongbuk Institute of Science and Technology (DGIST), Daegu 42988, Republic of Korea
babies are born yearly, in which 7% of those die due to preterm birth complications and lack of adequate newborn care. Also, in most countries, the premature birth rate is increasing [2, 3]. Primitive reflexes, a group of inborn behaviors found in normal early development, are one of the most freque
Data Loading...