Cardiac MRI is safe in patients with pacemakers and defibrillators

  • PDF / 152,401 Bytes
  • 2 Pages / 595.28 x 793.7 pts Page_size
  • 47 Downloads / 224 Views

DOWNLOAD

REPORT


ORAL PRESENTATION

Open Access

Cardiac MRI is safe in patients with pacemakers and defibrillators Alex Baher1*, Santhisri Kodali1, Jiaqiong Xu2, William A Zoghbi2, Miguel Quinones2, Miguel Valderrábano2, Dipan J Shah2 From 15th Annual SCMR Scientific Sessions Orlando, FL, USA. 2-5 February 2012 Background MRI has generally been contraindicated in patients (pts) with pacemakers (PM) or defibrillators (ICD) due to concern of a variety of potential complications. Several recent series have suggested potential safety of MRI in pts with cardiac devices, but most excluded cardiac or thoracic imaging. Additionally there is very limited data on safety of CMR in pts with devices, especially in pts with ICD. Methods Pts with devices underwent CMR on a 1.5-T Siemens Avanto scanner utilizing a pre-specified protocol that involved device interrogation and programming prior to, and immediately after CMR to assess lead impedance, battery voltage, and pacing capture/sensing thresholds. Devices in PM dependent pts (intrinsic HR < 60 bpm) were reprogrammed to asynchronous pacing mode, whereas devices in non-PM dependent pts were reprogrammed to pacing function deactivated. All underwent rhythm, pulse oximetry, and visual monitoring during CMR. All pts were scheduled for follow up device interrogation to evaluate for later changes in device parameters. For hospitalized pts troponin levels were measured prior to and 8 hours after CMR.

and there were no malignant arrhythmias or device malfunctions noted. Follow up device interrogation was performed after 68 (86%) of the exams at 13.2 ± 20.8 days after CMR. No device complications were noted during the follow up period. Table 1 and Table 2 summarize the changes in device parameters. We detected no significant rise in troponin levels with CMR (0.046 ± 0.200 and 0.037 ± 0.132, p = 0.596 for pts with PM; and value

Figure 1 Device parameters for PM and ICD measures before and immediately after CMR.

Results Seventy nine CMR exams were performed in 64 pts [26 men, age 64.7±14.2 years, 34 with PM, and 30 with ICD]. Of these, 47 exams [24 with PM, and 23 with ICD] were on inpatients who had troponin levels drawn before and after CMR. All pts underwent CMR without complications, specifically none complained of heating, 1

Figure 2 Device parameters for PM and ICD measures at follow up compared with the pre-test values.

Internal Medicine, The Methodist Hospital, Houston, TX, USA Full list of author information is available at the end of the article © 2012 Baher et al; licensee BioMed Central Ltd. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Baher et al. Journal of Cardiovascular Magnetic Resonance 2012, 14(Suppl 1):O102 http://www.jcmr-online.com/content/14/S1/O102

Page 2 of 2

0.099 ± 0.263 and 0.108 ± 0.247, p = 0.434 for pts with ICD).

Conclusions Patients with