Effect of crown to implant ratio and implantoplasty on the fracture resistance of narrow dental implants with marginal b

  • PDF / 1,357,846 Bytes
  • 10 Pages / 595.276 x 790.866 pts Page_size
  • 26 Downloads / 128 Views

DOWNLOAD

REPORT


EARCH ARTICLE

Open Access

Effect of crown to implant ratio and implantoplasty on the fracture resistance of narrow dental implants with marginal bone loss: an in vitro study Bruno Leitão‑Almeida1*  , Octavi Camps‑Font2, André Correia1, Javier Mir‑Mari2, Rui Figueiredo2 and Eduard Valmaseda‑Castellón2

Abstract  Background:  Peri-implantitis is a biological complication that affects soft and hard tissues around dental implants. Implantoplasty (IP) polishes the exposed implant surface, to decontaminate it and make it less prone to bacterial colonization. This study investigates whether a higher clinical crown-to-implant-ratio (CIR) reduces implant fracture resistance and whether implants are more fracture-prone after IP in the presence of 50% of bone loss. Methods:  Forty-eight narrow platform (3.5 mm) 15 mm long titanium dental implants with a rough surface and hexagonal external connection were placed in standardized bone-like resin casts leaving 7.5 mm exposed. Half were selected for IP. The IP and control groups were each divided into 3 subgroups with different clinical CIRs (2:1, 2.5:1 and 3:1). The implant wall width measurements were calculated using the software ImageJ v.1.51 through the analysis of plain x-ray examination of all the samples using standardized mounts. A fracture test was performed and scanning electron microscopy was used to evaluate maximum compression force ­(Fmax) and implant fractures. Results:  IP significantly reduced the implant wall width (P