Feasibility of hybrid TomoHelical- and TomoDirect-based volumetric gradient matching technique for total body irradiatio

  • PDF / 3,544,398 Bytes
  • 11 Pages / 595.276 x 790.866 pts Page_size
  • 89 Downloads / 151 Views

DOWNLOAD

REPORT


RESEARCH

Open Access

Feasibility of hybrid TomoHelical- and TomoDirect-based volumetric gradient matching technique for total body irradiation Chae-Seon Hong1, Min-Joo Kim1, Jihun Kim1, Kyung Hwan Chang1, Kwangwoo Park1, Dong Wook Kim1, Min Cheol Han1, Hong In Yoon1, Jin Sung Kim1* and Ho Lee2*

Abstract Background: Tomotherapy-based total body irradiation (TBI) is performed using the head-first position (HFP) and feet-first position (FFP) due to treatment length exceeding the 135 cm limit. To reduce the dosimetric variation at the match lines, we propose and verify a volumetric gradient matching technique (VGMT) by combining TomoHelical (TH) and TomoDirect (TD) modes. Methods: Two planning CT image sets were acquired with HFP and FFP using 15 × 55 × 18 cm3 of solid water phantom. Planning target volume (PTV) was divided into upper, lower, and gradient volumes. The junction comprised 2-cm thick five and seven gradient volumes (5-GVs and 7-GVs) to create a dose distribution with a gentle slope. TH-IMRT and TD-IMRT plans were generated with 5-GVs and 7-GVs. The setup error in the calculated dose was assessed by shifting dose distribution of the FFP plan by 5, 10, 15, and 20 mm in the longitudinal direction and comparing it with the original. Doses for 95% (D95) and 5% of the PTV (D5) were calculated for all simulated setup error plans. Absolute dose measurements were performed using an ionization chamber in the junction. Results: The TH&TD plan produced a linear gradient in junction volume, comparable to that of the TH&TH plan. D5 of the PTV was 110% of the prescribed dose when the FFP plan was shifted 0.7 cm and 1.2 cm in the superior direction for 5-GVs and 7-GVs. D95 of the PTV decreased to < 90% of the prescribed dose when the FF plan was shifted 1.1 cm and 1.3 cm in the inferior direction for 5-GVs and 7-GVs. The absolute measured dose showed a good correlation with the calculated dose in the gradient junction volume. The average percent difference (±SD) in all measured points was − 0.7 ± 1.6%, and the average dose variations between depths was − 0.18 ± 1.07%. Conclusion: VGMT can create a linear dose gradient across the junction area in both TH&TH and TH&TD and can minimize the dose sensitivity to longitudinal setup errors in tomotherapy-based TBI. Keywords: Total body irradiation, Tomotherapy, TomoHelical, TomoDirect, Field junction, Dose gradient, Setup error

* Correspondence: [email protected]; [email protected] 1 Department of Radiation Oncology, Yonsei Cancer Center, Yonsei University College of Medicine, 50-1 Yonsei-ro, Seodaemoon-gu, Seoul 03722, South Korea 2 Department of Radiation Oncology, Gangnam Severance Hospital, Yonsei University College of Medicine, 211 Eonju-ro, Gangnam-gu, Seoul 06273, South Korea © The Author(s). 2019 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit t