Performance assessment of the ALBIRA II pre-clinical SPECT S102 system for 99m Tc imaging

  • PDF / 1,432,205 Bytes
  • 10 Pages / 595.276 x 790.866 pts Page_size
  • 84 Downloads / 195 Views

DOWNLOAD

REPORT


ORIGINAL ARTICLE

Performance assessment of the ALBIRA II pre‑clinical SPECT S102 system for 99mTc imaging Ali Asgar Attarwala1 · Deni Hardiansyah1,2   · Chiara Romanó1 · Luis David Jiménez‑Franco1,3 · Mareike Roscher4,5 · Björn Wängler4 · Gerhard Glatting1,6 Received: 22 April 2020 / Accepted: 29 October 2020 © The Japanese Society of Nuclear Medicine 2020

Abstract Objective  The performance characteristics of the SPECT sub-system S102 of the ALBIRA II PET/SPECT/CT are analyzed for the 80 mm field of view (FOV) to evaluate the potential in-vivo imaging in rats, based on measurements of the system response for the commonly used Technetium-99 m (99mTc) in small animal imaging. Methods  The ALBIRA II tri-modal µPET/SPECT/CT pre-clinical system (Bruker BioSpin, Ettlingen, Germany) was used. The SPECT modality is made up of two opposite gamma cameras (Version S102) with Sodium doped Cesium Iodide (CsI(Na)) single continuous crystal detectors coupled to position-sensitive photomultipliers (PSPMTs). Imaging was performed with the NEMA NU-4 image quality phantom (Data Spectrum Corporation, Durham, USA). Measurements were performed with a starting activity concentration of 4.76 MBq/mL 99mTc. An energy window of 20% at 140 keV was selected in this study. The system offers a 20 mm, 40 mm, 60 mm and an 80 mm field of view (FOV) and in this study the 80 mm FOV was used for all the acquisitions. The data were reconstructed with an ordered subset expectation maximization (OSEM) algorithm. Sensitivity, spatial resolution, count rate linearity, convergence of the algorithm and the recovery coefficients (RC) were analyzed. All analyses were performed with PMOD and MATLAB software. Results  The sensitivities measured at the center of the 80 mm FOV with the point source were 23.1 ± 0.3 cps/MBq (single pinhole SPH) and 105.6 ± 5.5 cps/MBq (multi pinhole MPH). The values for the axial, tangential and radial full width at half maximum (FWHM) were 2.51, 2.54, and 2.55 mm with SPH and 2.35, 2.44 and 2.32 mm with MPH, respectively. The corresponding RC values for the 5 mm, 4 mm, 3 mm and 2 mm rods were 0.60 ± 0.28, 0.61 ± 0.24, 0.29 ± 0.11 and 0.20 ± 0.06 with SPH and 0.56 ± 0.20, 0.50 ± 0.18, 0.38 ± 0.09 and 0.23 ± 0.06 with MPH. To obtain quantitative imaging data, the image reconstructions should be performed with 12 iterations. Conclusion  The ALBIRA II preclinical SPECT sub-system S102 has a favorable sensitivity and spatial resolution for the 80 mm FOV setting for both the SPH and MPH configurations and is a valuable tool for small animal imaging. Keywords  SPECT · ALBIRA · Pre-clinical imaging

Ali Asgar Attarwala and Deni Hardiansyah contributed equally to this work. * Deni Hardiansyah [email protected] 1



Medical Radiation Physics/Radiation Protection, Universitätsmedizin Mannheim, Medical Faculty Mannheim, Heidelberg University, 68167 Mannheim, Germany

2

Medical Physics and Biophysics Research Group, Physics Department, Faculty of Mathematics and Natural Sciences (FMIPA), Universitas Indonesia, Depok 16424, Indonesia