Prospective assessment of plate-haptic rotationally asymmetric multifocal toric intraocular lens with near addition of + 1.5 diopters Tetsuro Oshika1* , Kazuno Negishi2, Toru Noda3, Hiroyuki Arai4, Mikio Inamura5, Yasushi Inoue6, Teruyuki Miyoshi7, Yoshifumi Fujita8, Kazunori Miyata9 and Yumi Hasegawa1

Abstract Background: To prospectively evaluate surgical results following implantation of rotationally asymmetric, platehaptic, refractive segmented multifocal toric intraocular lenses (IOLs) with near addition of + 1.5 diopters (D) (Lentis Comfort LS-313 MF15T, Oculentis GmbH). Methods: In 59 eyes of 41 patients, ocular examinations were conducted before and 1 day, 1 week, 1, 3, and 6 months after surgery. Uncorrected (UDVA) and corrected (CDVA) distance visual acuity, uncorrected (UIVA) and distance-corrected (DCIVA) intermediate visual acuity at 70 cm, and uncorrected (UNVA) and distance-corrected (DCNVA) near visual acuity at 30 cm were tested. A defocus curve was drawn, and the degree of disturbing photic phenomena were questioned. Results: The IOL showed excellent rotational stability; the average absolute rotation was 1.66 ± 1.17 degrees from 1 day 1 to 6 months postoperatively, and 98.1 and 100% of eyes yielded rotation of less than 5 and 10 degrees, respectively. Postoperative distance and intermediate visual acuity were highly satisfactory; UDVA, CDVA, UIVA, and DCIVA were about 20/20, 20/16, 20/25, 20/25, respectively. Near visual acuity was suboptimal; UNVA and DCNVA were at approximately 20/60. The defocus curve analysis showed that 20/25 and 20/40 uncorrected visual acuity was attained at as close as 60 and 40 cm, respectively. Contrast sensitivity was within a normal range, and subjective photic phenomena were minimum. Conclusions: The refractive segmented, rotationally asymmetric multifocal toric IOLs with + 1.5 D near addition showed superb rotational stability and highly satisfactory distance and intermediate vision. Contrast sensitivity was high and incidence of photic symptoms was very low. Trial registration: This study was registered at JAPIC Clinical Trials Information, ID: JapicCTI-183,877, https://www. clinicaltrials.jp/cti-user/trial/Search.jsp (February 5, 2018). Keywords: Intraocular lens, Multifocal lens, Toric lens, Rotationally asymmetric lens, Refractive segmented lens

* Correspondence: [email protected] 1 Department of Ophthalmology, Faculty of Medicine, University of Tsukuba, 1-1-1 Tennoudai, Tsukuba, Ibaraki 305-8575, Japan Full list of author information is available at the end of the article © The Author(s). 2020 Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's