Continuous sweep versus discrete step protocols for studying effects of wearable robot assistance magnitude

  • PDF / 1,260,345 Bytes
  • 13 Pages / 595.276 x 790.866 pts Page_size
  • 23 Downloads / 186 Views

DOWNLOAD

REPORT


RESEARCH

Open Access

Continuous sweep versus discrete step protocols for studying effects of wearable robot assistance magnitude Philippe Malcolm1,2,3* , Denise Martineli Rossi1,2,4, Christopher Siviy1,2, Sangjun Lee1,2, Brendan Thomas Quinlivan1,2, Martin Grimmer5 and Conor J. Walsh1,2

Abstract Background: Different groups developed wearable robots for walking assistance, but there is still a need for methods to quickly tune actuation parameters for each robot and population or sometimes even for individual users. Protocols where parameters are held constant for multiple minutes have traditionally been used for evaluating responses to parameter changes such as metabolic rate or walking symmetry. However, these discrete protocols are time-consuming. Recently, protocols have been proposed where a parameter is changed in a continuous way. The aim of the present study was to compare effects of continuously varying assistance magnitude with a soft exosuit against discrete step conditions. Methods: Seven participants walked on a treadmill wearing a soft exosuit that assists plantarflexion and hip flexion. In Continuous-up, peak exosuit ankle moment linearly increased from approximately 0 to 38% of biological moment over 10 min. Continuous-down was the opposite. In Discrete, participants underwent five periods of 5 min with steady peak moment levels distributed over the same range as Continuous-up and Continuous-down. We calculated metabolic rate for the entire Continuous-up and Continuous-down conditions and the last 2 min of each Discrete force level. We compared kinematics, kinetics and metabolic rate between conditions by curve fitting versus peak moment. Results: Reduction in metabolic rate compared to Powered-off was smaller in Continuous-up than in Continuousdown at most peak moment levels, due to physiological dynamics causing metabolic measurements in Continuousup and Continuous-down to lag behind the values expected during steady-state testing. When evaluating the average slope of metabolic reduction over the entire peak moment range there was no significant difference between Continuous-down and Discrete. Attempting to correct the lag in metabolics by taking the average of Continuous-up and Continuous-down removed all significant differences versus Discrete. For kinematic and kinetic parameters, there were no differences between all conditions. Conclusions: The finding that there were no differences in biomechanical parameters between all conditions suggests that biomechanical parameters can be recorded with the shortest protocol condition (i.e. single Continuous directions). The shorter time and higher resolution data of continuous sweep protocols hold promise for the future study of human interaction with wearable robots. Keywords: Exosuit, Protocol, Parameter sweep, Metabolic, Kinematics

* Correspondence: [email protected] 1 John A. Paulson School of Engineering and Applied Sciences, Harvard University, Cambridge, MA 02138, USA 2 Wyss Institute for Biologically Inspired Engineering, Harvard Universit