Cardiac MRI with concurrent physiological monitoring using MRI-compatible 12-lead ECG

  • PDF / 483,461 Bytes
  • 2 Pages / 595.28 x 793.7 pts Page_size
  • 28 Downloads / 156 Views

DOWNLOAD

REPORT


POSTER PRESENTATION

Open Access

Cardiac MRI with concurrent physiological monitoring using MRI-compatible 12-lead ECG Zion Tse1*, Charles Dumoulin2, Gari Clifford3, Julien Oster3, Michael Jerosch-Herold1, Raymond Kwong1, William Stevenson1, Ehud J Schmidt1 From 15th Annual SCMR Scientific Sessions Orlando, FL, USA. 2-5 February 2012 Background High fidelity 12-lead ECG is important for physiological monitoring during cardiovascular interventions. A dominant ECG R-wave is essential for synchronizing cardiac MRI. Obtaining the real ECG in MRI is challenging due to a superimposed Magneto-Hydro-Dynamic (MHD) voltage (VMHD) [1], & strong induced voltages from MRI switched gradients. Detecting acute ischemia by ST segment [2] is difficult due to VMHD peaks that occur during this period. We previously [1] presented (A) an adaptive MHD filtering procedure, based on 3 ECG training sets, & (B) R-wave detection based on 3-D ECG multichannel analysis. We extended our solution with (C) an electronic switching circuitry that blocks ECG transmission during Gradient Ramps and RadioFrequency Transmission (GR&RF), providing; (1) 12lead diagnostic-quality ECG free of MHD & GR&RF, (2) beat-to-beat stroke volumes (SV) estimated from VMHD, (3) accurately-gated cardiac MR images. Methods MR-compatible 12-lead ECG system (Fig.1) has 10 leads attached to patient, which transmit ECGs to the penetration panel on coaxial cables equipped with ferrite filters. Outside the room, an electronic circuit prevents ECGs from reaching the GE Cardiolab-IT digital ECGrecording system during GR&RF periods. Cardiolab streams out high-fidelity ECGs to a 64-bit computer, in which real-time MHD filtering & QRS detection are implemented. System outputs are real ECG, SV & Rwave triggers for scanner gating. System was validated on; 2 Atrial Fibrillation (AF), 1 Premature Ventricular Contraction patient, & 5 healthy subjects, including an

exercising athlete. Imaging was performed at 1.5T & 3T, with 12-lead ECG quality and electrode temperatures recorded. Before imaging, 20-sec breath-held ECGs were taken at 3 positions for MHD filter training (Fig.1 (2)). After imaging, the derived real ECGs were compared to ECGs measured periodically outside MRI for validation. Fig.1 (4) shows an improved R-wave detection using a 3-D ECG representation, consisting of a time, a voltage axes & a channels axis (ECG channels V1-V6).

Results AF ECG processing during a GRE scan (Fig.1) (3): (a) Raw ECG V6 is dominated by GR&RF noise, which is removed (b) by the switching circuit, leaving real ECG + VMHD. VMHD is removed (c) using adaptive filters & QRS detection, also providing (d) SV estimation (irregular due to changes in ventricular-filling). In (c) S-T segment is preserved for ischemia monitoring. 3-D QRS Rwave detection is demonstrated (Fig.1) (4): (a) Sinus rhythm 3-D QRS shape is distinguishable from (b) 3D MHD shape, (c) even when MHD voltage dominates. 3D-QRS (Fig.1(5)) correctly detected the QRS in ECGs acquired at 1.5T & 3T in all 8 subjects.