Analyzing medical device connectivity and its effect on cyber security in german hospitals
- PDF / 1,464,091 Bytes
- 15 Pages / 595 x 791 pts Page_size
- 89 Downloads / 186 Views
(2020) 20:246
RESEARCH ARTICLE
Open Access
Analyzing medical device connectivity and its effect on cyber security in german hospitals Markus Willing1†*
, Christian Dresen2† , Uwe Haverkamp1 and Sebastian Schinzel2
Abstract Background: Modern healthcare devices can be connected to computer networks and many western healthcare institutions run those devices in networks. At the same time, cyber attacks are on the rise and there is evidence that cybercriminals do not spare critical infrastructure such as major hospitals, even if they endanger patients. Intuitively, the more and closer connected healthcare devices are to public networks, the higher the risk of getting attacked. Methods: To asses the current connectivity status of healthcare devices, we surveyed the field of German hospitals and especially University Medical Center UMCs. Results: The results show a strong correlation between the networking degree and the number of medical devices. The average number of medical devices is 25.150, with a median of networked medical devices of 3.600. Actual key users of networked medical devices are the departments Radiology, Intensive Care, Radio-Oncology RO, Nuclear Medicine NUC, and Anaesthesiology in the group of UMCs. In the next five years, the usage of networked medical devices will increase significantly in the departments of Surgery, Intensive Care, and Radiology. We detected a strong correlation between the degree of connectivity and the likelihood of being attacked. The survey answers regarding the cyber security status reveal a lack of security basics in some of the inquired hospitals. We did discover successful attacks in hospitals with separated or subsidiary departments. A fusion of competencies on an organizational level facilitates the right behavior here. Most hospitals rated themselves predominantly positively in the self-assessment but also stated the usefulness of IT security insurance. Conclusions: Concluding our results, hospitals are already facing the consequences of omitted measures within their growing pool of medical devices. Continuously relying on historically grown structures without adaption and trusting manufactures to solve vectors is a critical behavior that could seriously endanger patients. Keywords: Networked medical devices, IT-security, Critical infrastructure, Germany
*Correspondence: [email protected] † Markus Willing and Christian Dresen contributed equally to this work. 1 University of Münster, Germany Full list of author information is available at the end of the article © The Author(s). 2020 Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless
Data Loading...