Image-based dosimetry for 225 Ac-PSMA-I&T therapy using quantitative SPECT
- PDF / 303,115 Bytes
- 2 Pages / 595.276 x 790.866 pts Page_size
- 51 Downloads / 214 Views
IMAGE OF THE MONTH
Image-based dosimetry for 225Ac-PSMA-I&T therapy using quantitative SPECT A. Gosewisch 1 & M. Schleske 1 & F. J. Gildehaus 1 & I. Berg 1 & L. Kaiser 1 & J. Brosch 1 & P. Bartenstein 1 & A. Todica 1 & H. Ilhan 1 & G. Böning 1 Received: 13 June 2020 / Accepted: 31 August 2020 # The Author(s) 2020
Targeted alpha therapy (TAT) using 225Ac-PSMA ligands is a promising therapy option for advanced metastatic castration-resistant prostate cancer (mCRPC) [1]. The 225Ac decay chain shows a noticeable gamma emission (440 keV, 25.9%; 218 keV, 11.4%). However, recommended low therapeutic activities (4–8 MBq) limit the clinical applicability of SPECT [2], although initial attempts for 225Ac imaging exist [3, 4]. Particularly quantitative SPECT is a vital tool to assess dosimetry and therapy response. While the 218-keV-peak is characterized by a lower branching ratio and a higher scatter fraction, SPECT imaging of high-energy gammas such as 440 keV causes a complex detector point spread function (PSF) [5]. In this study, we would like to demonstrate the general feasibility of image-based dosimetry for 225Ac radionuclide therapy using quantitative 225Ac SPECT. For a mCRPC patient (65 years), imaging of the abdomen was performed 24 h p. i. of 8.1 MBq 225Ac-PSMA-I&T on a Siemens Symbia Intevo T16 SPECT/CT (440 keV (width, 20%), lower adjacent window (width, 10%),
This article is part of the Topical Collection on Image of the month. * H. Ilhan [email protected] 1
Department of Nuclear Medicine, University Hospital, LMU Munich, Munich, Germany
HEGP collimator, 16 projections/head, 128 × 128 pixel, 210 s/projection). Reconstruction was carried out via a MAP algorithm (30i1s) [6], including CT-based attenuation and dual-energy-window scatter correction and a simulated distance-dependent 2D PSF model (SIMIND). Final absorbed dose assessment was performed by combining the single 225Ac image with the effective half-life information determined from a previous 177Lu-PSMAI&T imaging sequence [7]. This resulted in an absorbed dose of 0.18 and 0.17 SvRBE = 5/MBq for the left and right kidney, respectively, compared with 0.27 and 0.24 Gy/GBq for the preceding 177Lu cycle (6.2 GBq). A comparison with the pre-therapy 18 F-PSMA-I&T PET/CT demonstrates that 225Ac SPECT imaging for this patient was able to locate a small lesion in the right hip. The 225Ac-absorbed dose was determined as 0.26 SvRBE = 5/MBq, compared with 0.35 Gy/GBq for 177LuPSMA-I&T. Our analysis demonstrates the feasibility of dosimetry for 225 Ac-PSMA-I&T, which provides further insights into theranostic approaches using TAT in mCRPC patients.
Eur J Nucl Med Mol Imaging
Funding Open Access funding enabled and organized by Projekt DEAL.