In vivo comparison of braided (Accero) and laser-cut intracranial stents (Acclino, Credo): evaluation of vessel response

  • PDF / 1,421,163 Bytes
  • 9 Pages / 595.276 x 790.866 pts Page_size
  • 32 Downloads / 133 Views

DOWNLOAD

REPORT


BIOCOMPATIBILITY STUDIES Original Research

In vivo comparison of braided (Accero) and laser-cut intracranial stents (Acclino, Credo): evaluation of vessel responses at subacute and mid-term follow-up in a rabbit model Mühl-Benninghaus Ruben 1 Tomori Toshiki1 Krajewski Stefanie2 Dietrich Philipp1 Simgen Andreas1 Yilmaz Umut1 Brochhausen Christoph3 Kießling Mara3 Reith Wolfgang1 Cattaneo Giorgio4 ●

















1234567890();,:

1234567890();,:

Received: 1 March 2020 / Accepted: 27 October 2020 / Published online: 3 December 2020 © The Author(s) 2020

Abstract This study aimed to investigate in vivo two stent technologies, with particular emphasis on thrombogenicity and inflammatory vessel remodeling processes. The micro-stents tested in this study were developed for intracranial aneurysm treatment. In our study twelve, New Zealand white rabbits were divided into two groups: 18 laser-cut stents (LCS) and 18 braided stents (BS) were impanated without admiration of antiplatelet medication. Three stents were implanted into each animal in the common carotid artery, subclavian artery, and abdominal aorta. Digital subtraction angiography was performed before and after stent implantation and at follow-up for the visualization of occurring In-stent thromboembolism or stenosis. The Stents were explanted for histopathological examination at two different timepoints, after 3 and 28 days. Angiographically neither in-stent thrombosis nor stenosis for both groups was seen. There was a progressive increase in the vessel diameter, which was more pronounced for BS than for LCS. We detected a higher number of thrombi adherent to the foreign material on day 3 for BS. On day 3, the neointima was absent, whereas the complete formation observed was on day 28. There was no significant difference between both groups regarding the thickness of the neointima. The in vivo model of our study enabled the evaluation of blood and vessel reactions for two different stent technologies. Differences in vessel dimension and tissue around the stents were observed on day 28. Histological analysis on day 3 enabled the assessment of thrombotic reactions, representing an important complementary result in long-term studies. Graphical Abstract

Supplementary information The online version of this article (https:// doi.org/10.1007/s10856-020-06460-z) contains supplementary material, which is available to authorized users. * Mühl-Benninghaus Ruben [email protected]

3

Department of Pathology, University of Regensburg, Regensburg, Germany

1

Department of Neuroradiology, Saarland University Hospital, Homburg/Saar, Germany

4

Institute for Biomedical Engineering, University of Stuttgart, Stuttgart, Germany

2

Department of Thoracic, Cardiac and Vascular Surgery, University Hospital, Tuebingen, Germany

131

Page 2 of 9

Journal of Materials Science: Materials in Medicine (2020) 31:131

Keywords In vivo comparison of braided and laser-cut stents Thromboembolism Neointima ●

Abbreviations AA abdominal aorta BS braided stent CCA common carotid arte