Retention force of polyetheretherketone and cobalt-chrome-molybdenum removable dental prosthesis clasps after artificial

  • PDF / 926,392 Bytes
  • 9 Pages / 595.276 x 790.866 pts Page_size
  • 21 Downloads / 205 Views

DOWNLOAD

REPORT


ORIGINAL ARTICLE

Retention force of polyetheretherketone and cobalt-chrome-molybdenum removable dental prosthesis clasps after artificial aging Felicitas Mayinger 1 Bogna Stawarczyk 1

&

Danka Micovic 1 & Andreas Schleich 1 & Malgorzata Roos 2 & Marlis Eichberger 1 &

Received: 29 April 2020 / Accepted: 9 October 2020 # The Author(s) 2020

Abstract Objectives To examine the retention force of removable dental prosthesis (RDP) clasps made from polyetheretherketone (PEEK) and cobalt-chrome-molybdenum (CoCrMo, control group) after storage in water and artificial aging. Materials and methods For each material, 15 Bonwill clasps with retentive buccal and reciprocal lingual arms situated between the second pre- and first molar were manufactured by milling (Dentokeep [PEEKmilled1], NT digital implant technology; breCAM BioHPP Blank [PEEKmilled2], bredent), pressing (BioHPP Granulat for 2 press [PEEKpressed], bredent), or casting (remanium GM 800+ [CoCrMo], Dentaurum); N = 60, n = 15/subgroup. A total of 50 retention force measurements were performed for each specimen per aging level (initial; after storage [30 days, 37 °C] and 10,000 thermal cycles; after storage [60 days, 37 °C] and 20,000 thermal cycles) in a pull-off test. Data were statistically analyzed using one-way ANOVA, post hoc Scheffé and mixed models (p < 0.05). Results Initial, PEEKpressed (80.2 ± 35.2) and PEEKmilled1 (98.9 ± 40.3) presented the lowest results, while PEEKmilled2 (170.2 ± 51.8) showed the highest values. After artificial aging, the highest retention force was observed for the control group (131.4 ± 56.8). The influence of artificial aging was significantly higher for PEEK-based materials. While PEEKmilled2 and PEEKpressed showed an initial decline in retention force, all other groups presented no impact or an increase in retention force over a repetitive insertion and removal of the clasps. Conclusions Within the tested PEEK materials, PEEKmilled2 presented superior results than PEEKpressed. Although CoCrMo showed higher values after artificial aging, all materials exhibited sufficient retention to recommend usage under clinical conditions. Clinical relevance As RDPs are still employed for a wide range of indications, esthetic alternatives to conventional CoCrMo clasps are sought. Keywords PEEK . Cobalt-chrome-molybdenum . Clasp . Removable dental prosthesis . Retention force

Introduction Felicitas Mayinger and Danka Micovic contributed equally to this work. * Felicitas Mayinger [email protected] 1

Department of Prosthetic Dentistry, University Hospital, LMU Munich, Goethestraße 70, 80336 Munich, Germany

2

Department of Biostatistics, Epidemiology, Biostatistics and Prevention Institute, University of Zurich, Hirschengraben 84, 8001 Zurich, Switzerland

Removable dental prostheses (RDPs) are commonly used to treat patients with large or multiple edentulous areas. Indications furthermore include the replacement of missing teeth in patients with severely damaged periodontal tissue, an excessive loss of alveolar bone