The diagnostic accuracy of quantitative CMR perfusion imaging may not be the same for all coronary arteries
- PDF / 197,208 Bytes
- 2 Pages / 595.28 x 793.7 pts Page_size
- 19 Downloads / 169 Views
POSTER PRESENTATION
Open Access
The diagnostic accuracy of quantitative CMR perfusion imaging may not be the same for all coronary arteries Geraint Morton*, Amedeo Chiribiri, Masaki Ishida, Andreas Schuster, Shazia T Hussain, Eike Nagel From 15th Annual SCMR Scientific Sessions Orlando, FL, USA. 2-5 February 2012 Background Absolute quantification of myocardial perfusion with cardiac magnetic resonance (CMR) is an emerging technique and its diagnostic accuracy remains to be fully established. Furthermore it is not known whether there are any differences in the accuracy of coronary artery disease (CAD) detection between coronary arteries. Methods 41 patients with known or suspected CAD underwent 1.5T perfusion imaging prior to coronary angiography. Perfusion imaging
k-t balanced Turbo Field Echo sequence, shortest TE (range 1.35-1.54ms), shortest TR (range 2.64-3.12ms), 50°flip angle; 90° prepulse, 100ms prepulse delay and typical acquired resolution 1.7x1.9x10mm. 3 short axis slices were acquired every heartbeat. Stress imaging preceded rest by 14±2 minutes. A dual bolus (0.01mmol/kg then 0.1mmol/kg 20 seconds later) of contrast agent was used. Blinded experts performed all data analysis. Mean segmental perfusion values were obtained using dedicated prototype Philips ViewForum software and Fermi deconvolution. Each segment was assigned to a perfusion territory according to standard definitions. Myocardial perfusion reserve (MPR) was defined as stress perfusion divided by rest perfusion. The mean of the 2 lowest MPR for each perfusion territory was used for analysis.
X-ray angiography
Quantitative coronary angiography (QCA) was performed on all arteries ≥2mm and a mean diameter stenosis ≥70% regarded as significant. The accuracy of MPR to detect significant CAD as defined by QCA was determined by receiver-operating characteristic (ROC) curves.
Results Mean age was 63±9. 32 (78%) participants were male, 13 (32%) diabetic and 30 (73%) hypertensive. 5 (12%) had previous myocardial infarction. LVEF was 63±12%. 25 patients had significant CAD involving 7 left anterior descending (LAD) arteries, 12 circumflex arteries (3 dominant) and 13 right coronary arteries (RCA; 12 dominant). Rest perfusion was 1.03±0.3ml/min/g in stenotic and 1.06±0.33ml/min/g in remote territories. Corresponding stress perfusion values were 1.54±0.34ml/min/g and 1.94 ±0.59ml/min/g and MPR 1.31±0.3 and 1.7±0.42. Area under the ROC curve for MPR to detect significant CAD was 0.83 (0.74-0.92). MPR ≤1.45 detected CAD with 82% sensitivity and 81% specificity. ROC curves for each individual coronary artery are shown in figure 1. Area under the ROC curve, sensitivity and specificity were highest for the RCA and lowest for the circumflex (table 1). Area under the ROC curve was significantly lower for the circumflex artery compared with the RCA (p=0.04) and non-significantly lower (p=0.1) compared with the LAD. There was no significant difference between the LAD and RCA ROC curves (p=0.6).
King’s College London, London, UK © 2012 Morton et al; licens