Towards clinical application of 7T TOF angiography
- PDF / 183,450 Bytes
- 2 Pages / 595.28 x 793.7 pts Page_size
- 42 Downloads / 165 Views
WORKSHOP PRESENTATION
Open Access
Towards clinical application of 7T TOF angiography Sebastian Schmitter*, Edward J Auerbach, Gregor Adriany, Kamil Ugurbil, Pierre-Francois Van de Moortele From 15th Annual SCMR Scientific Sessions Orlando, FL, USA. 2-5 February 2012 Background As shown at 7T Time-of-Flight (TOF) angiography significantly benefits from ultra high field (UHF) [1,2], because of shorter T1 relaxation constant and increased SNR. However, significant challenges have to be overcome before using 7T TOF in clinical applications: I) RF induced specific absorption rate (SAR) substantially increases at UHF, preventing the use of standard RF pulses (venous Saturation (SAT), Magnetization Transfer (MT)) to improve TOF contrast. II) Severe transmit B1 (B1+) heterogeneity at 7T impairs contrast homogeneity. Methods In this work we approached both challenges with complementary strategies: 1) SAR reduction with VERSE [3,4] for excitation and travelling venous SAT pulses. The max amplitude of VERSE RF pulses was set to a fraction EXC (or SAT) of the max amplitude of initial excitation (or SAT) pulse. 2) Sparse application of MT [5] (10% of the centered k-space lines in phase and slab direction [6]). 3) Improved homogeneity of RF excitation profile with B1+ shimming (16 transmit channels), using fast multi channel B1+ calibration [7]. The standard deviation divided by the mean (std/mean) of B1+ was minimized over the center slices of 3 TOF slabs using nonlinear optimization algorithms. Results 1+2) Fig.1 shows maximum intensity projections (MIPs) for a) the original sequence with EXC =100% (no VERSE), without SAT and without MT, and b) our modified sequence with EXC=50%, with venous SAT (SAT=25%) and sparse MT. Quantitative results: vessel University of Minnesota, Center for Magnetic Resonance Research, Minneapolis, MN, USA
to tissue contrast improved by >50%, venous signal suppressed by ~90%. Strong fat signal suppression observed within the excitation slab results from the large spatial chemical shift of VERSE-SAT pulse. SAR values were, for a) and b), 17% and 69% of maximum SAR limit, respectively. By comparison, using standard excitation, SAT and MT RF pulses (no VERSE/no sparse RF) would amount 319% of SAR limit. 3) Homogeneity strongly improved with B1+ shimming (Fig.2), with std/mean reduced by a factor 2.2. Smaller vessels in the brain periphery show higher contrast or become visible (see arrows). In general, however, spatial homogeneity optimization reduces B1+ efficiency, thus increases SAR. For example, for the excitation pulse SAR with B1 shimming was 4.7 times the value without B1 shimming, however this pulse represents only a fraction of SAR.
Conclusions We demonstrate, with appropriate RF pulse design, substantial gains in contrast and B1+ homogeneity for TOF angiography at 7T. These results strongly support the potential of expanding non-contrast enhanced angiography at 7T towards clinical investigations. Funding ACKNOWLEDGEMENT: DFG: SCHM 2677/1, NIH P41 RR008079, P30 NS057091, R01 EB000331,
Data Loading...