Enhanced Biocompatibility of GPC by MeV Ion Bombardment 1 2 1 1 3 R. Zimmerman , I. Gurhan , S. Sarkisov , C. Muntele1, and D. Ila and M. Rodrigues 1 Center for Irradiation of Materials, Alabama A&M University, Normal, AL 2 Ege University Faculty of Engineering, Izmir, Turkey 3 University of Sao Paulo, Ribeirao Preto SP Brazil
I ABSTRACT
9
Glassy Polymeric Carbon (GPC) is completely biocompatible and is widely used
10
as a material for artificial heart valves and in other biomedical applications. Although it
11
is ideally suited for fluid flow in the blood stream, collagenous tissue that normally forms
12
around the moving parts of a GPC heart valve sometimes loses adhesion and creates
13
embolisms downstream. We have shown that moderate fluence of MeV ions, especially
14
oxygen ions, increases the surface roughness of GPC on a scale appropriate for enhancing
15
tissue adhesion. Silver ion implantation is shown to inhibit cell growth on GPC. Ion
16
bombardment also increases the surface hardness of GPC, already an extremely hard
17
material. In vitro biocompatibility tests have been carried out with model cell lines to
18
demonstrate that MeV ion bombardment can favorably influence the surface of GPC for
19
biomedical applications.
20 21 22 23 24 25 26 27 28
Key Words Carbon biomaterials, Carbon, Tissue adhesion Corresponding Author Robert Zimmerman Alabama A&M University Research Institute P. O. Box 1447 Normal, ALABAMA 35672 TEL 256372 5854 FAX 256 372 5868 [email protected]
K3.10.2
28
II INTRODUCTION
29
Glassy Polymeric Carbon (GPC) is an outstanding and versatile biomaterial [1-5]
30
owing to its electrical, mechanical and chemical properties, which include exceptional
31
biocompatibility. The density (1.45 g/cm3) of GPC is significantly lower than that of
32
graphite from which one may deduce a relative pore volume of about 35% although the
33
material is among the most impermeable known. Its structure has intertwined graphene
34
planes that enclose voids.
35
The strength, durability and low density make GPC a favored material for the
36
manufacture of replacement heart valves. However the low adhesion at the interface with
37
biological tissue [6] has the potential of creating an embolism when endothelial tissue
38
that naturally encapsulates the implant is released into the blood stream. Greater tissue
39
adhesion can be achieved by making the surface rough. The texture dimensions and
40
chemical affinity of the surface should be adequate to allow strong adherence of tissue on
41
the material surface. Alternatively, surface treatment by an appropriate ion implantation
42
may be expected to inhibit cell growth or give the cells no adherence at all.
43 44
We report the use of oxygen ion bombardment to increase the surface roughness, to enhance cell adhesion, and implantation of silver ions that inhibit cell growth.
45 46
III EXPERIMENTAL METHODS