Comparison of the Conox (qCON) and Sedline (PSI) depth of anaesthesia indices to predict the hypnotic effect during desf

  • PDF / 1,742,634 Bytes
  • 8 Pages / 595.276 x 790.866 pts Page_size
  • 60 Downloads / 235 Views

DOWNLOAD

REPORT


ORIGINAL RESEARCH

Comparison of the Conox (qCON) and Sedline (PSI) depth of anaesthesia indices to predict the hypnotic effect during desflurane general anaesthesia with ketamine Catherine Christenson1 · Pablo Martinez‑Vazquez2,3   · Max Breidenstein1 · Borzoo Farhang1 · Jackson Mathews1 · Umberto Melia2 · Erik Weber Jensen2,4 · Donald Mathews1 Received: 14 February 2020 / Accepted: 9 November 2020 © Springer Nature B.V. 2020

Abstract Comparison of two depth of anesthesia indices, qCON (Conox) and PSI (Sedline), during desflurane sedation and their sensitivity to random ketamine boluses in patients undergoing routine surgery. The performance of desflurane and ketamine on both indices was analyzed for 11 patients, and the ketamine sensitivity was compared with another group of 11 patients under sevoflurane and propofol. The MOAA/S was used to determine sedation level and pain. Different boluses of ketamine ranging from 10 to 30 mg where randomly administered in both groups and the effect on the indexes were measured after 4 min. The indices were recorded during the whole surgery, and their correlations with the desflurane concentration and the discrimination between awake and anesthetized states were evaluated with the prediction probability statistic (Pk). The Pk values, mean (se), discriminating between awake and anesthetized states were 0.974(0.016) for the qCON and 0.962(0.0123) for the PSI, while the 1-Pk statistic for the qCON and the PSI with respect to the desflurane concentration were 0.927(0.016) and 0.918(0.018), respectively, with no statistically significant differences. The agreement between both depth of hypnosis parameters was assessed under the Bland-Altman plot and the Spearman correlation, rs = 0.57(p