Diagnostic analysis of new quantitative parameters of low-dose dynamic myocardial perfusion imaging with CZT SPECT in th

  • PDF / 1,198,914 Bytes
  • 12 Pages / 595.276 x 790.866 pts Page_size
  • 21 Downloads / 184 Views

DOWNLOAD

REPORT


REVIEW PAPER

Diagnostic analysis of new quantitative parameters of low‑dose dynamic myocardial perfusion imaging with CZT SPECT in the detection of suspected or known coronary artery disease Zekun Pang1 · Jiao Wang1 · Shuai Li1 · Yue Chen1 · Xiaojie Wang1 · Jianming Li1  Received: 15 April 2020 / Accepted: 3 August 2020 © The Author(s) 2020

Abstract The goal of this study is to explore and evaluate the diagnostic values of myocardial blood flow (MBF), myocardial flow reserve (MFR) and relative flow reserve (RFR) obtained with low-dose dynamic CZT SPECT for patients with suspected or known coronary artery disease (CAD). Fifty-seven consecutive patients who underwent low-dose dynamic CZT SPECT and CAG were enrolled. MBF, MFR and RFR were calculated on the vessel level with dedicated quantitative software, and the difference and correlation of each parameter was compared according to the reference standard of stenosis ≥ 50% or ≥ 75% on CAG, respectively. ROC curves were made by stress MBF (sMBF), rest MBF (rMBF), MFR and RFR. The optimal cut-off values and corresponding diagnostic efficacy were obtained and compared with each other. Results indicated that when stenosis ≥ 50% or ≥ 75% on CAG was used as the reference standard at the vessel level, there was no statistically significant difference in rMBF between the negative group and the positive group (P > 0.05), and the sMBF and MFR in positive groups were significantly lower than that in the negative group (all P 